Metabolite annotation

Iterative exclusion-list MS/MS acquisition

Use precursors fragmented in earlier injections to guide subsequent DDA runs toward additional ions.

Journal of the American Society for Mass Spectrometry · 2017 · 10.1007/s13361-017-1608-0

LC-MSUntargeted metabolomicsMetabolite annotation

What does this strategy enable?

The IE-Omics workflow automates exclusion-list preparation between repeat injections. This record uses its 2017 metabolomics methods paper as a reproducible implementation; iterative exclusion has earlier proteomics origins. The defining relationship is measured fragmentation history informing the next acquisition, rather than precomputing survey-based schedules.

Transferable workflow

Workflow

  1. 1

    Acquire the initial DDA run

    Begin with top-N fragmentation of a representative sample.

  2. 2

    Convert acquired spectra

    Export the measured run to MS2 format with msconvert.

  3. 3

    Generate the next exclusion list

    IE-Omics groups fragmented precursors by mass and retention time and can flag recurrent background ions.

  4. 4

    Import the list and repeat acquisition

    Load the CSV into the instrument method before the next injection; the reported workflow is not real-time control.

  5. 5

    Process and annotate spectra

    The study uses MZmine 2 feature detection with LipidMatch, alongside LipidSearch annotation.

Connected resources

Tools used in this strategy

Strategy profile

Related objectives

Molecular discovery

Components

Data acquisition, Preprocessing and feature extraction, Annotation or identification

Biological context

Broadly applicable

Ion mobility

Not specified

Scope and considerations

Designed for

  • Extending fragmentation coverage when repeated LC-MS/MS injections are feasible.

Important requirements

  • Sufficient sample and instrument time for repeated injections.
  • Compatible instrument exclusion-list import; the documented implementation targets Thermo instruments.

Considerations

  • Check mass and retention-time tolerances and chromatographic reproducibility with the MS facility.
  • Additional spectra do not guarantee identification or equal gains across sample types.
  • This is not a replacement for quantitative study design or quality control.

Reusable resources

✓ Code available✓ Published method available

Publication

Koelmel JP, Kroeger NM, et al. Journal of the American Society for Mass Spectrometry 28, 908-917 (2017).